| Regardless of how you use specialty gas
| |
| | at the bottom of the traceability
|
| mixtures to calibrate a
| |
| | hierarchy. A GMIS introduces more
|
| laboratoryinstrument or monitor, the most
| |
| | uncertainty because more steps are
|
| important thing you can do to ensure the
| |
| | involved and because NIST is not involved
|
| accuracy of your measurements is to
| |
| | in either their analysis or
|
| ensure the accuracy of your gas mixture.
| |
| | certification. The gas supplier is
|
| Gas mixture accuracy, in turn, is largely
| |
| | responsible for assigning the
|
| dependent upon its analytical
| |
| | concentration and accuracy, based solely
|
| traceability to a reference material. For
| |
| | on their analysis.
|
| analytical traceability to exist, there
| |
| | Gas mixtures that are produced using
|
| must be an unbroken link of comparison
| |
| | GMISs do not provide direct
|
| between the measured value of the gas
| |
| | NIST-traceability, and so may not be
|
| mixture and a reference that is
| |
| | suitable for some applications. For
|
| recognized in the national or
| |
| | example, the EPA requires that a
|
| international measurement system as being
| |
| | Continuous Emission Monitor be calibrated
|
| a true value. Thus, analytical
| |
| | using a protocol that is at least as
|
| traceability is the link between the
| |
| | accurate to within ±2%. Use of a
|
| analyzed concentration and the true
| |
| | protocol that was prepared using a GMIS
|
| concentration of a gas mixture. The
| |
| | may not guarantee such accuracy.
|
| National Institute of Standards and
| |
| | When selecting any specialty gas mixture,
|
| Technology (NIST) provides Standard
| |
| | be sure you know what type of standard
|
| Reference Materials (SRMs) for this
| |
| | the mixture you are purchasing will be
|
| purpose. Another such metrological
| |
| | analyzed against. To ensure accuracy,
|
| organization, the Netherlands Measurement
| |
| | insist on traceability to an SRM, PRM or
|
| Institute (NMi), provides Primary
| |
| | NTRM. Also remember that, as in the case
|
| Reference Materials (PRMs).
| |
| | of emission monitoring, traceability and
|
| Analytical traceability is achieved by
| |
| | subsequent accuracy may be governed by
|
| laboratory analysis of a mixture using an
| |
| | law.
|
| instrument that has been directly
| |
| | You can ensure the integrity of your
|
| calibrated using either an SRM or a PRM.
| |
| | laboratory calibrations by obtaining gas
|
| A mixture analyzed in this manner is said
| |
| | mixtures from reliable companies such as
|
| to be "directly traceable." SRMs and PRMs
| |
| | Scott Specialty Gas, which has been the
|
| are at the top of the traceability
| |
| | leading supplier of precursor standards
|
| hierarchy and add the least amount of
| |
| | for NIST SRMs for more than 25 years,
|
| uncertainty to a mixture analysis because
| |
| | Praxair, the second largest supplier of
|
| they are individually analyzed and
| |
| | industrial and specialty gases to
|
| certified by NIST or NMi laboratories.
| |
| | chemical companies worldwide, and Airgas,
|
| Since 1995, NIST and NMi have maintained
| |
| | which offers an extensive line of high
|
| an equivalence agreement in those cases
| |
| | purity specialty gases as well as a full
|
| where comparative analyses have been
| |
| | range of specialty gas equipment
|
| completed by the two labs. Because of
| |
| | specifically designed for use with
|
| this NIST/NMi equivalency, traceability
| |
| | highest purity gases and high tolerance
|
| to a PRM is considered the legal
| |
| | gas mixtures. Your dedication to accurate
|
| equivalent of an SRM.
| |
| | calibrations will be rewarded many times
|
| Direct analytical traceability can also
| |
| | over in the precision of your
|
| be established using what are known as
| |
| | measurements, and thus the quality of
|
| NIST Traceable Reference Materials
| |
| | your products, which can only make your
|
| (NTRMs). NTRMs are produced by gas
| |
| | customers smile.
|
| suppliers and then batch analyzed and
| |
| | For more information about EPA protocol
|
| audited by NIST against the same internal
| |
| | gas accuracy, download a copy of Accuracy
|
| standards as SRMs. NTRM values are
| |
| | Counts from
|
| assigned by NIST and offer nearly the
| |
| | This article is provided by Scott
|
| same accuracy confidence level as SRMs
| |
| | Specialty Gases. Scott Specialty Gases, a
|
| and PRMs in most cases.
| |
| | leading global manufacturer of specialty
|
| The high cost of producing SRMs and PRMs
| |
| | gases located in Plumsteadville, PA. More
|
| make it economical for some specialty gas
| |
| | information on the company can be found
|
| suppliers to blend intermediate standards
| |
| | at
|
| called Gas Manufacturer Intermediate
| |
| | This article is copyrighted by Scott
|
| Standards (GMISs). Though they are
| |
| | Gases.
|
| analyzed against existing SRMs, they are
| |
| |
|